Head, Probe and Sample Preparation
Initial Preparation for Contact AFM Imaging
Rev. B
MultiMode SPM Instruction Manual
73
5.1.2 Load the Sample
Remove Head and Load Sample
1. If the head is not already removed, do so now by unfastening the retaining springs on either
side and unplugging the head’s micro-D connector.
2. Gently lift the head off and set aside. This will expose the top of the scanner tube.
3. Mount the sample puck with the calibration standard atop the scanner tube. An internal
magnet supplied with most units holds the puck down.
Figure 5.1b
MultiMode Base with Scanner Mounted on Support Ring
Reinstall the Head
•
With the sample in place, remount the head by gently lowering it over the scanner tube
while checking for clearance. The top of the sample should protrude no more than a few
millimeters above the top of the head’s X-Y translation stage.
•
Secure both retaining springs and plug the head’s connector into the support ring
(see
Top of scanner tube.
Retaining springs
Place sample puck here.