
Force Imaging
Force Calibration (Contact Mode AFM)
Rev. B
MultiMode SPM Instruction Manual
191
Motor
The
Motor
menu allows you to withdraw or manually control the tip position using the stepper
motor. Selecting
Step Motor
opens a dialog box containing the following buttons:
•
Tip Up:
This command moves the tip up by the
SPM step size
displayed inside the
window.
•
Tip Down:
This command moves the tip down by the
SPM step size
displayed inside
the window.
11.4 Force Calibration (Contact Mode AFM)
11.4.1 Obtaining a Good Force Curve
Figure 11.4a
Typical Force Calibration Curve
To minimize or calculate the contact force between the tip and sample, obtain a good force curve
which shows the typical features displayed in the example curve in
. However, the
force curve rarely looks “typical” immediately after invoking
Force Calibration
mode. This
section discusses general approach adjustments to improve force curves obtained after engaging the
microscope.
The basic approach to obtaining a good force curve entails adjusting the Z motion of the piezo
relative to the sample (with the
Z scan start
and
Ramp size
parameters) and shifting the graph
(with the
Setpoint
parameter) so the pull-off point of the tip displays on the graph. In general, use
the following steps to obtain a good force curve:
1.
Engage
and verify the tip is tracking the surface. Then switch to
Force Calibration
mode.
2. Maximize the
Deflection Limit
parameter in the
Feedback
Controls
panel.
(Z Scan start -Ramp size)
Setpoint
Slope = Volts of Deflection/nanometers (or volts) of piezo travel
+
Cantilever Off Surface
(Z Scan start)
Z Piezo Voltage
1
2
3
4
5
6
V
csmin
Cantile
ver De
fl
ection
V
oltage