22
When the temperature alarm setting key (F4)
is pushed in the operation mode, the “Alarm
Range” page shown on the left appears. The
desired alarm value can be adjusted by
pushing value increase (F1) and decrease (F2)
keys. Enter key (F4) is pushed to save the
value and returned to the working screen. If
you want to exit this page without any
change, push backspace key (F3).
PERIODIC MAINTENANCE AND CLEANING
PERIODIC MAINTENANCE
The dust and grease settled on the condenser fin reduce the area of the heat exchange and thus
efficiency of freezer decreases. Therefore, the condenser filter should be cleaned once in a
month. To access condenser filter, pull out the front panel of the compressor compartment (See
Figure 5). Take out the filter and clean the filter with compressed air. Place the filter again on
the front panel of the compressor compartment.
Figure 5
Icing on the door gasket may be observed. In this case, the door will not be closed strictly and
the icing in the chamber will start. Therefore, the cooling may not be effective. Please check
regularly whether there is any icing at the door gasket.
If excessive icing is observed, it is suggested to turn off the freezer and wait for 12
hours.
Clean the chamber with surface disinfectant (%70 Alcohol, %2 glutaraldehyde). Do not use
solutions including chlorine for cleaning.
CLEANING
Clean the refrigerator and at room temperature after disconnecting the power cable.
Clean the refrigerator with a damp cloth to remove dirt and dust.
Use liquid detergent to remove tough dirt.
Take precautions while handling chemical cleaners.
Check the external condition of the refrigerator regularly and ensure any rust spots that may
develop are removed.
Ensure that the temperature of chamber is reached to ambient temperature before
cleaning.
Ensure that there is no ice on the shelves. Do not inscribe layer of ice if there is ice on
shelves.