10 Advanced Parameters and Commands
29 / 41
Example:
Integration of 32 images simultaneously to the output of the actual integration image
and copy as a reference image
A
into the correction data memory (a=1, b=6, c=1) =>
yyxxxw = „011101“ binary = 0x1D. As soon as the integration is terminated, one still
has to wait for at least two image cycles, before the copy procedure is ended.
>
H=1D[CR]
...wait for at least 36 image cycles (34 + 2 frames jitter buffer, 0.9 seconds at 40
images / sec.) or poll the state of bit group
d
with
H[CR]
...
>
H=0[CR]
10.7
Baud Rate (s)
The Baud Rate of the RS232 interface is determined by use of register
s
:
The bits of the parameter value are divided into three different sized bit groups
a
,
e
and
x.
Each bit
group has a special function, being independent of the other ones:
•
a
: Controls the configuration of the serial port.
•
e
: Controls the echo mode.
•
x
: Controls the baud rate of the serial port.
F E D C B A 9 8 7 6 5 4 3 2 1 0
- - - - - - - - e a a - x x x x
s 0
Operation Mode
Integrator / Image Store
rw rw
rw
rw rw rw rw
Bit group
Value (hex)
Explanation
x
O
110 Baud
1
300 Baud
2
600 Baud
3
1200 Baud
4
2400 Baud
5
4800 Baud
6
9600 Baud
7
19200 Baud
8
38400 Baud
9
57600 Baud
A
115200 Baud
a
00
RS232 via 15 pin SUB-D only.
01
Additionally activate second channel via Pleora device port serial 0.
10
Additionally activate second channel via Pleora device port bulk 0 (MODE:UART).
11
Reserved.
e
0
Each character received from the host is echoed back to it.
1
No Echo.
Example:
Set the factory default (115200 Baud and activate serial interface via Camera Link /
Gigabit Ethernet).
>
s=2A[CR]
VDS Vosskühler GmbH
IRC-320