17
Caution
Once the patient information being entered, the photos and videos taken will be appended with the
patient information until the patient information is cleared.
Note
The image showed in the image zone will be cleared after the patient ID is modified.
6.1.2.
White Balance Adjustment
This adjustment procedure is used to display the correct image color on the monitor. Be sure to always adjust the
white balance in the following cases:
* Before observation.
*After exchange of the endoscope
*Restart the Digital Video Monitor
*When any abnormality can be seen on the color of the image even if white balance adjustment has been
completed.
Warning
When adjusting the white balance of the endoscope, use a white object such as a piece of gauze without
bringing it in contact with the endoscope. Contact of the endoscope with a non-sterilized object may result in
cross-contamination.
Caution
When adjusting the white balance, make sure the light built in the distal end is on and take care not to expose
the distal end of the endoscope to external light. Otherwise, it may cause an incorrect white balance adjustment.
1.
The adjustment procedure is used to display the correct image color on the monitor. Be sure to always adjust the
white balance before observation or when any abnormality can be seen on the color of the image even if white
balance adjustment has been completed.
2.
Hold the endoscope stable to avoid wash-out of the monitor image, monitoring a white object such as a piece of
gauze in such a way that it does not contact the endoscope,, contact of the endoscope with a non-sterilized object
may result in cross-contamination. Vathin Medical does not provide any accessories (e.g. white cap) for white