12
To avoid the risk of electric shock, only the appropriately approved medical electrical equipment are
allowed to be connected with the product.
The product can only be connected to a medical display screen that has been qualified as per IEC
60601-1.
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation
.
Do not use the product in case of any damage to it or any unacceptable item in the functional inspection.
Otherwise, operator/patient injury may occur.
1.
Prepare Digital Video Monitor, power cable, power adapter, cables and device used together. Carefully
check the product and all the components for damage (and without wear). Never use the product in case of
any damage.
2.
Prepare and inspect the endoscope according to its instruction manual.
5.2.
Installation of Digital Video Monitor
Warning
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective
earth. Electric shock and/or fire may be caused if the product is not properly grounded.
All the system components shall be powered off before connecting them. Only appropriate cables can be
used for the product;Otherwise, the product may be damaged or malfunction.
Caution
Always place the power cable in the place where it is unlikely to be stepped on. Do not place any objects
on the power cable.
1.
Open the bracket (No. 12
in Section 4.3 Structure and Function
) of the product and place it on a solid
surface. Do not place the Digital Video Monitor in a place which is difficult to operate the disconnection
device.
2.
Connect the power cable to the power adapter, and connect the power adapter to the power jack ( No.8
in Section 4.3 Structure and Function
) of the Digital Video Monitor.
3.
Connect the power cable to a supply mains with protective earth. The power supply should be stable. In
case of severe fluctuations in the power supply, a power regulator or UPS power supply should be used for