B. Installing the Emergency Stop Switch
Install the
Emergency Stop Switch
in the power cable line.
Install this switch so that it is easy to reach in the emergency case but can’t be
pushed by mistake.
The switch shall be a type of mistake-proof.
The switch is not supplied.]
The switch shall be installed at a height of 1.2 to 1.5 meters (47 to 60”).
1.
The cable sizes: N, L and PE≥ 12 AWG (3 x 4 mm
2
).
2.
The cable to
Emergency Stop Switch
shall be the same size as the power cable itself.
3.
Install the socket connector terminal for the 2nd protective ground wire.