14.3 Electrical Specifications
Item
Description
Power supply voltage
100 - 240 V ~
Frequency
50 / 60 Hz
Power rating
2.0 kVA
Accuracy
Tube
Voltage (kVp) ± 10 %, Tube Current (mA) ±
20 %, Exposure Time (s)
±
(5 % + 50 ms)
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ±10 %.
Mode of operation: Continuous operation with intermittent loading - Needs waiting
time (at least 60 times the exposure time) before the next exposure begins.
Column operation time: Max. 2 min. On / 18 min. Off (Ratio 1:9)