6.
PANO Geometry Calibration
6.1. PANO Geometry Calibration Standard
•
To perform the PANO Geometry Calibration, PANO BALLPHANTOM should be
prepared in advance.
•
During the PANO BALLPHANTOM Geometry operation, the measurements
must be checked by using the Raw File.
•
PANO BALLPHANTOM RIGHT/LEFT PIN DISTANCE ERROR
-
within
±
30 pixels
•
PANO BALLPHANTOM CENTER PIN POSITION
-
Within 1574 ±10 pixel
•
PANO BALLPHANTOM CENTER BALL SIZE
-
54 ~ 55 pixels
CENTER PIN
CENTER BALL SIZE
Right/Left Pin Distance