6.2.
CEPH Geometry Calibration
6.2.1.
Lateral Mode Start Position Setting
1. Identify the starting position using the CEPH standard phantom.
2. Attach 0.8T of Cu in front of the collimator and set the exposure condition to 80kVp 10mA.
3. Attach CEPH standard phantom to the ear rod, click the console software as shown in the
picture below, and capture the Normal LAT Mode.
5. Choose the
Ready button
.
1. Choose the
CEPH
mode
2. Choose the
High resolution
.
3. Choose the
Lateral
.
4. Choose the
Confirm
button.