15
TriScroll 300 Dry Scroll Vacuum Pump
DRAFT 2
/
10
/
0
5
19. Lightly grease the new 2-269 O-ring and
install it around the lip on the inboard
housing.
20. Insert the tip seals in to the grooves on the
outboard housing.
21. Cut the seal to the correct length at the end
of each groove. Leave a gap of 1/8" (3mm)
from the outer end to allow for thermal
growth.
TriScroll 300 Reassembly (continued)
O-ring