1-1
SECTION I
INSTALLATION
1-1
GENERAL
Upon receipt, inspect the container for evidence of damage in shipment. Do not discard any evidence of
rough handling. Factory packing provides for maximum protection during shipment. However, the pump
should be inspected immediately and any damage should be reported to the carrier without delay.
Before unpacking and installing the pump, the user should thoroughly familiarize himself wiith the
contents of this instruction manual, particularly the physical and operating specifications listed in the
following paragraphs, in order to gain a better understanding of the operating principles, limitations,
correct application of the diffusion pump, the hazards involved, and what to do and what not to do in
order to avoid these hazards.
1-2
OPERATING SPECIFICATIONS
Model Part
Number
NHS-35 Diffusion Pump
240 V
480 V
380/415 V
400 V
440 V
with ASA flanges & standard cold cap F1730308
F1730309
F1730326 F1730362
F1730364
with ASA flanges and halo baffle
F1730318
F1730319
F1730336 F1730363
F1730365
with ISO flanges & standard cold cap
L5924308
L5924309
L5924326
–
–
with ISO flanges and halo baffle
L5924318
L5924319
L5924336
–
–
Optimum Operating Range
6 x 10
-4
torr to < 5 x 10
-8
torr
(8 x 10
-2
Pa to < 7 x 10
-6
Pa)
(8 x 10
-4
mbar to < 7 x 10
-8
mbar)
Maximum Pumping Speed
50,000 liters/sec for air
(See Figure 1-1)
62,500 liters/sec for helium
Maximum Throughput
35 torr-liters/sec at 10
-2
torr
4655 Pa - liters/sec @ 1.33 Pa
46.5 mbars - liters/sec @ 0.013 mbar
Maximum Forepressure
No Load – 0.55 torr (73.2 Pa) (0.73 mbar)
Full Load – 0.40 torr (53.2 Pa) (0.53 mbar)
Backstreaming Rate at Pump Inlet
0.0005 mg/cm
2
/min (with cold cap)
Power Required (approximately)
24 KW, 240/380/415/440/480V, 50/60 Hz, 3-phase
Warmup Time
60 hour
Cooldown Time (using quick cool)
180 minutes (60 minutes)
Fluid Charge
3 gallons
Cooling Water Requirements
4 gpm at 60°F to 80°F inlet temperatures
Backing Pump Size Recommended
300 cfm for optimum throughput
Summary of Contents for NHS-35
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...x...
Page 18: ......
Page 20: ......
Page 22: ...Installation 1 2 Figure 1 1 Speed and Throughput Curves NHS 35 Diffusion Pump...
Page 26: ...Installation 1 6 Figure 1 3 Section Drawing NHS 35 Diffusion Pump...
Page 30: ...Installation 1 10 Figure 1 5 240 400 440V Wiring Diagram NHS 35 Diffusion Pump...
Page 31: ...Installation 1 11 Figure 1 6 380 415V Wiring Diagram NHS 35 Diffusion Pump...
Page 32: ...Installation 1 12 Figure 1 7 480V Wiring Diagram NHS 35 Diffusion Pump...
Page 36: ...Installation 1 16...
Page 48: ......