iv
Hazard
Suggested Corrective Action
a. Explosion – Operation of the diffusion pump without prior and continuous evacuation (below 0.5
Torr (66.5 Pa)) or coolant and introducing a strong oxidizer (such as air) or explosive vapors or
powders or materials which may react with pumping fluids into a hot (above 300°F or 150°C) pump
can cause an explosion. Such an explosion can violently expel valves and other hardware, slam
open doors that are not designed for appropriate pressure relief, or burst other components of the
vacuum system. Serious injury or death may result from expelled parts, doors, shrapnel, and/or
shock waves.
b. All diffusion pumps are typically cleaned with acetone and alcohol. When combined with air,
oxygen, and other oxidizers, alcohol and most other solvents are very flammable and explosive.
Never permit any trace of these cleaners to remain in or on the pump. Always remove all traces of
alcohol and acetone and other cleaners with clean, dry, oil-free compressed air.
Three elements are required: fuel, oxidizer, and source of ignition. A combination of temperature and
pressure can be a source of ignition. Most diffusion pump fluids except mercury are fuels. Hydrocarbon
oils are more prone to oxidize and explode than synthetic silicone-based oil. Oxidizers can be air (a
strong oxidizer) from a leak or can be deliberately introduced in a process or can be inadvertently
admitted to the system by operator or process controller error. Oxygen and other strong oxidizers are
even more dangerous. Certain conditions of temperature and pressure can cause a combustible mix-
ture to explode. The larger the diffusion pump, the greater the risk of explosion and the greater the risk
of damage and/or injury. Never operate diffusion pumps larger than 10 inches in diameter with hydro-
carbon oils without a full safety analysis for the complete system and application. Never operate any dif-
fusion pump in the following ways because they increase the probability of an explosion.
Prohibited Action
Explosion-Causing Condition
Run pump with no cooling water
Overtemperature
Run pump with low level of pump fluid
Overtemperature
Run pump without proper backing or
Overpressure
holding pump
Run pump when not evacuated below
Overpressure
0.5 Torr (66.5 Pa)
Admit air to, or rough through,
Overpressure plus strong oxidizer
a pump with hot boiler
Open drain or fill plug while pump is
Overpressure plus strong oxidizer
under vacuum, especially when it is hot
Summary of Contents for NHS-35
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...x...
Page 18: ......
Page 20: ......
Page 22: ...Installation 1 2 Figure 1 1 Speed and Throughput Curves NHS 35 Diffusion Pump...
Page 26: ...Installation 1 6 Figure 1 3 Section Drawing NHS 35 Diffusion Pump...
Page 30: ...Installation 1 10 Figure 1 5 240 400 440V Wiring Diagram NHS 35 Diffusion Pump...
Page 31: ...Installation 1 11 Figure 1 6 380 415V Wiring Diagram NHS 35 Diffusion Pump...
Page 32: ...Installation 1 12 Figure 1 7 480V Wiring Diagram NHS 35 Diffusion Pump...
Page 36: ...Installation 1 16...
Page 48: ......