www.vareximaging.com
37
XRpad2 4343
U
SER
M
ANUAL
11.2
Monthly Inspection
Perform the following inspection at least once a month. If there is a problem, immediately
ask your establishment’s safety department to contact your dealer, distributor, or device
manufacturer.
11.3
Yearly Inspection
Perform the following inspection at least once a year. If there is any problem, immediately ask your
establishment’s safety department to contact your dealer, distributor, or device manufacturer.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Ge
nera
l
Execute the Dark Noise
and X-ray Uniformity tests
(refer to the
XRpad2 4343
Digital X-ray Detector
Reference Manual
).
Good/Bad Good/Bad Good/Bad If there are changes in
performance, acquire new
calibration files as described
in the
Digital Radiography
Software Manual
.
Contact your dealer,
distributor, or device
manufacturer if there is any
problem.
Make sure that the XRpad
IPU-2 is clean from dirt or
dust.
Good/Bad Good/Bad Good/Bad Use the instructions in
“Section 10.5.1.6, Cleaning
the Battery Pack” on page 22
for cleaning.
Check the manufacturing
date of all battery packs to
ensure that their age is not
five years or older.
Good/Bad Good/Bad Good/Bad Replace the battery pack with
a new one and recycle the old
one.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Ge
ner
a
l
Execute the Dark Noise,
X-ray Uniformity, Bad
Pixel, and Resolution tests
(refer to the
XRpad2 4343
Digital X-ray Detector
Reference Manual
).
Good/Bad Good/Bad Good/Bad If there are changes in
performance, acquire new
calibration files as described
in the
Digital Radiography
Software Manual
.
Contact your dealer,
distributor, or device
manufacturer if there is any
problem.
Version
01
2017-11-03