XRpad2 4343
36
www.vareximaging.com
U
SER
M
ANUAL
11.1.2
After Turning On the Power
11.1.3
After Turning Off the Power
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Ge
ner
a
l
Check that the wireless
connectivity or the AP
symbol ( )is shown in
the display if wireless mode
is used.
Good/Bad Good/Bad Good/Bad Connect the X-ray detector
and the Wi-Fi Access Point as
described in the
Access Point
Manual
.
Check that the LAN
connectivity symbol ( ) is
shown in the display if the
wired mode is used.
Good/Bad Good/Bad Good/Bad Connect the Gigabit Ethernet
cable and the tethered power
and communication cable
properly.
Check the battery charge
condition (Battery Half or
better (
).
Good/Bad Good/Bad Good/Bad Exchange the battery pack
with a charged one.
Check that the detector LED
is ON.
Good/Bad Good/Bad Good/Bad Set the Detector to
Exposure
Ready
as described in the
Digital Radiography Software
Manual
.
Perform test exposure as
described in the
Digital
Radiography Software
Manual
.
Good/Bad Good/Bad Good/Bad If any error messages appear,
follow the instructions in the
Digital Radiography Software
Manual
. If there is a problem,
contact your dealer,
distributor, or device
manufacturer.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
General
Check that the X-ray
detector is turned off
normally and that all LEDs
are OFF.
Good/Bad Good/Bad Good/Bad See
Off the X-ray Detector” on
page 32
for turning off the
X-ray detector.
Make sure that the X-ray
detector is clean and
disinfected.
Good/Bad Good/Bad Good/Bad See
and Disinfecting
Instructions” on page 38
for
cleaning the X-ray detector.