10.5.2.1 Cleaning the XRpad IPU-2
If the XRpad IPU-2 (Interface and Power Unit) surface is dirty or dusty, clean it with a cleaning
cloth dampened with ethanol or a diluted neutral detergent. If you are using a disinfectant other
than those specified, we recommend you consult a specialist for the procedure for disinfection.
Turn off the XRpad IPU-2, and disconnect the AC power cable, detector power, and detector
communication tethered cables before cleaning.
10.6
Minimum System Requirements
The following are the minimum requirements for the host computer that controls the X-ray
detector.
1
Gigabit Ethernet Infrastructure and a free Gigabit Ethernet Port or Wi-Fi Infrastructure.
2
Intel compatible Multi Core Processor (> 2 GHz).
3
RAM > 4 GB.
4
Windows 7 (32-bit/64-bit).
5
If a Firewall is used, make sure that it allows a connection to the detector.
6
Access Point:
a
WPA2 encryption support.
b
802.11n standard with 20 MHz channel bandwidth 36-48, 52-64, 100-116, 132-140,
149-165.
c
MIMO 3x3.
d
Complying with IEC 60601-1 or IEC 60950-1.
WARNING
When the Interface and Power Unit is going to be cleaned, be sure to
turn off the XRpad IPU-2, and unplug all cables. Never use thinner,
benzine, acetone, or other flammable cleaning agents. Ignoring this
warning may result in explosion, fire, or electric shock, which may
result in severe personal injury, death, or substantial product damage.