For Your Safety
WARNING
The detector is not designed to control X-ray dose. The system
integrator is responsible for controlling the X-ray radiation.
WARNING
Connection of the X-ray detector to the installed IT network that
includes other equipment could result in previously unidentified risk
to patients, operators or third parties. the responsible organization
should identify, analyze, evaluate and control these risk. Subsequent
changes to the IT-network could introduce also new risks and require
additional analysis.
Caution
Do not operate the X-ray detector in a location with the following
conditions.
•
Close to fluid or places where fluid is used
•
Close to heat sources, such as a heater
•
High temperature environment
•
High humidity environment
•
High condensation environment
•
Extreme cold environment
•
Dusty environment
•
Salty or sulphurous environment
•
Near a vibrating environment
•
Environment where there is insufficient air circulation to enable the X-ray
detector, and power supply to dissipate heat.
Ignoring this caution may result in personal injury or damage to
the product.
Caution
When using the X-ray detector in Automatic Exposure Detection
(AED) mode, the temperature may increase due to the heat generated
by the X-ray detector electronics to support the AED mode. Follow the
integration guide and instructions for the use of the X-ray detector in
AED mode. Ignoring this caution may result in personal injury or
damage to the product.
Caution
Monitor the temperature of the X-ray detector and allow the detector to
cool down when the temperature of the detector gets hot. Ignoring this
caution may result in personal injury or damage to the product.
Caution
An Image Quality Test or Customer Acceptance test should be
performed before the detector is used for the first time with patients.