1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
40
www.vareximaging.com
U
SER
M
ANUAL
10.5.4
Fluoroscopic Mode Gain Calibration Steps
Perform the following steps to complete fluoroscopic gain calibration.
1
Click the
desired mode
.
Figure 40
Mode Drop Down Menu
2
Click
Acquisition
from the menu bar and click on
Gain Calibration
.
WARNING
It is critical that the detector not be turned off or the Ethernet cable
removed during any calibration.
Caution
ViVA provides a convenience of running offset calibration as part of the
gain calibration process; however, API-driven gain calibrations do not
automatically run offset calibrations. OEM applications should be sure to
run offset calibration prior to gain calibration.