Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2019-01-23
31/56
www.vareximaging.com
Service screen 3
Show Logo
With this setting, the logo on the
Main screen
can be made invisible(Off) or visible(On).
The default setting is: On
Show Inclino
With this setting, the Inclino box on the
Main screen
can be made invisible(Off) or visible(On).
The default setting is: On
Software revisions are shown in the lower left corner:
Main:
contains the version for the main board.
Display:
contains the version for the display module.
If the Main version is not shown, either the display and main versions are not compatible or there is no
communication between the display module and the main board.
Warning:
Do not use the device in case there is no communication between display
module and the main board
Remark: The versions shown in the image above are not the actual versions for your device, this is
only an example!
X-ray beam Alignment
Centering the imaging modality to the X-ray beam
Warning:
The procedure described in this section requires that X-rays are produced by
the X-ray tube. Take adequate precautions to ensure that the human body is not exposed
to X-radiation, direct or indirect!
Please, consult the X-ray system manual to align the imaging modality to the X-ray beam.
Alternatively, the following instructions are provided
When the collimator is mounted on the X-ray system, fully open the shutters.
To check alignment of the imaging modality to the X-ray beam it is necessary to do the following:
•
Place a small metal tube approximately 20mm long and 8 mm in diameter in the center of the
imaging modality.
•
When X-rayed and the image shows a perfect circle, the focal spot is exactly centered.