Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2019-01-23
15/56
www.vareximaging.com
H
Bucky laser shutter to close the laser aperture
I
Accessory rails, for a maximum of 2 accessories with a width of 177 mm
J
Accessory lock, for locking the accessory
Prior to inserting accessories in the accessory rails check the fit and locking.
Hazard of falling accessory.
K
Touch screen (Optica 40 Series only)
Use of the measuring tape:
The type of measuring tape used is a standard measuring tape. The tape used in the
collimator starts with values that correspond to the focus/collimator lower edge distance;
maximum measurement with the tape is 2 m.
Forcing and/or extending the tape beyond this point could cause the following
inconveniences:
•
Breakage of the tape or,
•
Distortion of the tape or,
•
Impossibility of retracting the tape into its housing.
NOTE
: Do not extend the tape beyond the indication of 2.0 m
NOTE
: High speed release of the tape might damage the tape