© May 2022 Van Essen Instruments. All rights reserved.
www.vanessen.com
8
* The CTD-Diver is always in stand-by when not making a measurement. The power consumption
of the integrated battery is dependent on the temperature and usage.
If the CTD-Diver is used, stored or transported for extended periods of time under high temperature,
reduced, but this is not permanent. This is normal behavior for batteries.
Excessive programming, high frequency sampling and data reading will reduce the battery capacity.
** The accuracy of the clock is highly dependent on temperature. The clock is actively compensated
for temperature variations.
Environmental
Ingress protection
IP68, 10 years continuously submerged in water at 200 m
Transportation
Suitable for transportation by vehicles, ships and airplanes in the supplied packaging.
Resistance to vibration
In accordance with MIL-STD-810.
Mechanical shock test
In accordance with MIL-STD-810, for light-weight equipment
Temperature
-20 °C to 80 °C (affects battery life)
Temperature
Measurement range
-20 °C to 80 °C
Operating Temperature (OT)
0 °C to 50 °C
Accuracy (max)
± 0.2 °C
Accuracy (typical)
± 0.1 °C
Resolution
0.01 °C
Response time (90% of final value)
3 minutes (in water)