Hyperion Optical Sensors - Operating Manual:
MANUAL-360093108-14 | issue: 2.1
Page | 11
©
2023
–
Valeport Ltd
3.2.1
Mode C - Continuous
Mode C is the default setting for Hyperion.
Command example: #039;C02
In this mode the instrument will be set into a continuous output cycle of appropriately averaged and
filtered data.
Mode C02: (C zero 2) will output a reading every 2 seconds. This is the maximum synchronous
observation measurement period.
Mode C04: (C zero 4) will output a reading every 4 seconds but that will be a mean of 2, 2 second
synchronous observation measurement periods.
Setting Measurement period (sec) Synchronous sample interval (secs)
Number of samples
C01
1
1
1
C02
2
2
1
C04
4
2
2
C06
6
2
3
Cxx
xx
2
xx/2
3.2.2
Mode M - Continuous Measurement
Mode M will perform a measurement and data output at the rate specified up to 16Hz.
Command example: #039;M8
Filter settings are appropriately set for the update rate.
M1: performs continuous measurements at 1Hz
M2: performs continuous measurements at 2Hz
...
M16: performs continuous measurements at 16Hz
Setting Measurement period (sec) Synchronous sample interval (secs)
Number of samples
M1
1
1
1
M2
0.5
0.5
1
M4
0.25
0.25
1
M8
0.125
0.125
1
M16
0.0625
0.0625
1
3.2.3
S Mode
Single observation on run command (#028)
Command example: #039;S2
Setting Measurement period (sec) Synchronous sample interval (secs)
Number of samples
S1
1
1
1
S2
2
2
1
S4
4
2
2
S6
6
2
3
Sxx
xx
2
xx/2