![Vaisala WINDCAP WMT700 Series User Manual Download Page 173](http://html.mh-extra.com/html/vaisala/windcap-wmt700-series/windcap-wmt700-series_user-manual_849074173.webp)
12. Select
File > Properties
.
13. Select
Properties > Configure
.
14. In
Bits per second
, select
9600
.
15. Close the
Properties
window.
16. Select
Call > Call
.
Table 58 Restored Serial Port Settings
Parameter Name
Default Value
Description
com1_baudcom2_baud
4
4 = 9600 bauds per second
com1_data
com2_data
8
Number of data bits
com1_delay
com2_delay
20
Response delay in milliseconds
com2_interf
0
0 = RS-485
com1_parity
com2_parity
0
0 = None
com1_protocol
com2_protocol
0
WMT700 protocol
com1_stop
com2_stop
1
Number of stop bits
The serial port settings have now been restored to known values. You can write the
OPEN
command and start configuring WMT700.
More Information
‣
Chapter 7 – Troubleshooting
171
Summary of Contents for WINDCAP WMT700 Series
Page 1: ...M211095EN H User Guide Vaisala WINDCAP Ultrasonic Wind Sensor Series WMT700 ...
Page 32: ...WMT700 User Guide M211095EN H 30 ...
Page 100: ...WMT700 User Guide M211095EN H 98 ...
Page 168: ...4 Remove the verifier WMT700 User Guide M211095EN H 166 ...
Page 174: ...WMT700 User Guide M211095EN H 172 ...
Page 182: ...WMT700 User Guide M211095EN H 180 ...
Page 194: ...WMT700 User Guide M211095EN H 192 ...
Page 198: ...WMT700 User Guide M211095EN H 196 ...
Page 201: ...Appendix G Certificate Appendix G Certificate 199 ...
Page 202: ...WMT700 User Guide M211095EN H 200 ...
Page 208: ...WMT700 User Guide M211095EN H 206 ...
Page 209: ......
Page 210: ...www vaisala com ...