Chapter 5 ____________________________________________________Updating the Software
VAISALA _______________________________________________________________________ 37
CHAPTER 5
UPDATING THE SOFTWARE
This chapter contains instructions for software updating.
MRP111 Software Update
The update CD contains latest version of unit software for MRP111.
CAUTION
Power must be on all the time while the update is in progress.
Switching off the unit before the update is finished may cause serious
malfunction.
1.
Switch on the Sounding Workstation and the SPS311 and wait
until the system is ready for operation.
2.
Place the Setup CD into the CD-ROM drive of the Sounding
Workstation and click
Start
, select
Run
and click the
Browse
button.
3.
Locate the CD-ROM drive and go to folder <CD-ROM
drive>\MRP111, click the
update.exe
and click
Open
. Click
OK
.
4.
A command line window opens and the text "Enter the
hostname or the IP address of the updated Embedded PC" is
displayed. Type the IP address of the MRP111 and press
ENTER. The default IP address for the MRP111 is
192.168.0.10.
5.
For the question "Verify previous update?" Press
n
and ENTER
.