A.1.2 Configuration registers
Table 15 Modbus configuration data registers (writable)
Register
number
(decimal)
Address
(hexadecimal)
Register description
Data format
Unit / Valid range
Environmental compensation
769
0300
hex
Power-up value for
pressure compensation
32-bit float
hPa
Range: 450 … 1550
Init/default: 1013.25
771
0302
hex
Power-up value for
temperature
compensation
32-bit float
°C
Range: -10 … 70
Init/default: 25
773
0304
hex
Volatile pressure
compensation (value
cleared at probe reset)
32-bit float
hPa
Range 450 ... 1550
hPa
Init/default: 1013.25
775
0306
hex
Volatile temperature
compensation (value
cleared at probe reset)
32-bit float
°C
Range: -10 … 70
Init/default: 25
Function control
1281
0500
hex
Pressure compensation
mode selection
Enum
0
= Off
1
= Setpoint
2
= External
1282
0501
hex
Temperature
compensation mode
selection
Enum
0
= Off
1
= Setpoint
2
= Measured
3
= External
Communication
1537
0600
hex
Serial address
16-bit integer
Valid range 1 …
255
Default: 240
MGP260 User Guide
M212504EN-A
72
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:[email protected]
Summary of Contents for MGP260 Series
Page 8: ...MGP260 User Guide M212504EN A 6 Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 88: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 90: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 91: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Page 92: ...www vaisala com Tel 029 88186182 Web www xavery cn Email sxvery 163 com...