outputs a help text. The new system prompt includes the user's
previous input up to the first parameter requiring clarification. This
allows the user to complete the command with the correct parameter.
The help text contains explanations for only those parameters that
match the operator's input.
Table 10
User Level Commands
Command Description
CLOSE
Releases the port for message transmission.
DUMP
Prints out a snapshot of the current instrument system
situation and configuration.
ECHO
ECHO OFF
Disables input character echo. Set off when operating with
the RS-485 connection.
ECHO ON
Enables input character echo (default)
EVENTS
Gives the events logged by the system. It gives for each
event listed the following information:
- the time of occurrence backward from now by
years, months, days, hours, minutes, and
seconds
- the time of occurrence in operating seconds
-
the type ( S - single, B bad, or G - good) and
the event text is given
EVENTS
page
The list is given by separate pages. The page length is
configured by SET TERMINAL_LINES.
EVENTS
string
The event list is reduced to event which event text contains
string
.
EVENTS
string page
The reduced event list is given by separate pages.
HELP Lists
available
command.
HELP
command
Shows information about a specific command.
LEVEL 1
or
LEVEL 1 [password]
Changes the operation level to the advanced level.
The password can be given as a parameter. If it is not given
as a parameter, the user is asked for it.
The default password for the advanced level is LT31 (case
sensitive).
MEAS_SYNC
Resets the measurement sequence. The following data
message is measured starting from the moment all the
average values are erased. The next message is transmitted
in seconds (in2).
MENU
Starts the menu-driven operation. For a proper display, the
terminal application must be set to VT100 emulation. For
proper operation, the terminal height in lines used by the
terminal application must be configured to LT31 with the
command SET TERMINAL_LINES
number
(minimum 24).
Also, any 'line wrap'- and 'translate CR to CR/LF' -option
must be switched off for the terminal application in use.
MESSAGE
Displays the default message without transmission frames.
MESSAGE ALL
LT31 displays all messages.
MESSAGE
message_number
LT31 displays the selected message with transmission
frames.
NAME
Displays the device type, device name given by the user,
and the ID.
PARAMETERS
Prints configuration parameters
SET
SET DATA_PORT BAUD
number yes/no
Sets data port serial line transmission rate to 300 ... 19200.
Verification yes/no asked only if the command is given
through DATA PORT (default 9600).