8. Type
20
in the
Scale high end
field.
9. Type
3.6
in the
Error output level
field.
10. Type
5
in the
Clipping limit
field and
10
in the
Error limit
field.
11. Select
Save
to take your configuration into use.
5.2.1 Adjusting analog output channels
If the analog output of the transmitter appears to be inaccurate, you can test and adjust the
output level in the Insight PC software calibration menu. In analog output adjustment, the
output of the transmitter's analog output channels is forced to a fixed current or voltage level.
The test output can then be verified with a multimeter, and the analog output level adjusted as
necessary.
For instructions on testing and adjusting analog output, see
.
5.3 Clearing transmitter analog output
configuration
If you need to restore the transmitter to a state where no analog output configuration exists in
the transmitter, you can clear the current configuration from the Insight main menu.
When you remove the analog output configuration from the transmitter, the
transmitter automatically adapts the analog output settings of the next probe you
connect to it.
1. Connect the transmitter to the Insight PC software.
2. Select
to open the Insight main menu.
Indigo201 User Guide
M212842EN-B