![Vaisala Indigo 500 Series User Manual Download Page 31](http://html2.mh-extra.com/html/vaisala/indigo-500-series/indigo-500-series_user-manual_3387257031.webp)
Table 7 Connection terminals for probe 1 and probe 2
Terminal
Function
Wire color in Vaisala
cables
Notes
RS-485-
RS-485-
White
Max. wire size: 2.5 mm
2
(14 AWG)
RS-485+
RS-485+
Black
24 VOUT -
Power GND and
RS-485 common
Blue
24 VOUT +
Positive supply voltage to
probe
Brown
3.3.7 Verifying tightness of cable glands
To maintain enclosure tightness and provide strain relief to the cables, all cable glands must be
tightened. Unused cable glands must remain plugged.
1. Check every cable gland that is in use:
a. Pull on the cable to verify that the cable is securely held by the cable gland.
b. Tighten the cable gland if the cable moves.
2. Check that every unused cable gland is plugged and tightened.
3.4 Attaching probes
Attach only Vaisala Indigo compatible probes to the transmitter.
CAUTION!
The IP classification of probes is valid only when the probes are
attached to the probe connection cable.
CAUTION!
If you want to configure the probe settings, such as purge interval, you must do
that before attaching the probe to the transmitter. To configure the probe, you
can use the free Vaisala Insight PC software. For more information, see your
probe's user guide and visit
.
When the probe connection cables have been wired to the transmitter, you can attach and
detach probes both when the transmitter power is on and when it is off. If the power is on
while you attach or detach a probe, the transmitter shows a notification about the probe. If
you have configured the transmitter to use outputs, the transmitter also notifies you of the
state of the outputs.
Chapter 3 – Installation
29
Summary of Contents for Indigo 500 Series
Page 1: ...M212287EN A User Guide Indigo 500 Series Transmitters Indigo 520...
Page 8: ...Indigo 520 User Guide M212287EN A 6...
Page 80: ......
Page 81: ......
Page 82: ...www vaisala com...