Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 41
Reset Transmitter
When the
RESET
command is given, the transmitter will restart as if it
had just been powered on. All transmitter settings are retained. The serial
line mode of the transmitter will be set to the mode that was set using the
SMODE
command.
RESET
<cr>
Restore Factory Settings
Use the
FRESTORE
command to restore the factory settings to the
transmitter. All user settings will be lost.
FRESTORE
<cr>
ppm Conversion
MMT162 ppm Conversion for
Transformer Oils
Traditionally, moisture in transformer oil is measured by using ppm
units. The ppm output shows the average
mass concentration of water
in
oil.
The moisture and temperature transmitter MMT162 has an option for
ppm-output provided that this has been notified when placing the order
for the transmitter. Vaisala has this conversion readily available for
mineral transformer oils.