![Vaisala HUMICAP HMT360 SERIES User Manual Download Page 81](http://html.mh-extra.com/html/vaisala/humicap-hmt360-series/humicap-hmt360-series_user-manual_849332081.webp)
Chapter 9 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 79
0507-021
Figure 27
Accuracy over Temperature Range
Water Activity in Jet Fuel
Applications
Table 9
Water Activity in Jet Fuel Applications
Specifications
Property
Description / Value
Measurement range
0 ... 1 a
w
at -40 ... +180 °C (-
40 ... +356 °F)
Accuracy when calibrated against high
quality, certified humidity standards
salt solutions (ASTM E104-85)
± 0.01 (0 ... 0.9)
± 0.02 (0.9 ... 1.0)
± 0.02 (0 ... 0.9)
± 0.03 (0.9 ... 1.0)
Response time (90 %) at +20 °C in still oil
(stainless steel filter)
10 min
Humidity sensor
Vaisala HUMICAP
®
Calculated variables available
mass concentration of water
relative humidity of saturation
saturation temperature
Accuracy of the water content
better than ± 15 % of the
reading