USER'S GUIDE____________________________________________________________________
38 __________________________________________________________________ M211024EN-B
View User Adjustment Parameters
Use the
L
command to view the current user adjustment parameters and
trimmer adjustment. This command is useful for checking the currently
applied correction.
L
<cr>
The output values are as follows:
- Cp offset and gain: capacitance, calibrated using the
CRH
command
- RH trimmer offset and gain: correction by the trimmers
- RH offset and gain: calibrated using the
RHCAL
command
- T offset and gain: calibrated using the
CT
command
Example (shows default values, no user calibration done):
>l
Cp offset : 0
Cp gain : 1
RH offset trimmer : 0
RH gain trimmer : 1
RH offset : 0
RH gain : 1
T offset : 0
T gain : 1
>
Set User Adjustment Parameters
Use the
LI
command to set the user adjustment parameters directly. This
command is useful for restoring the desired adjustments, and removing
the effects of calibration without having to perform the factory reset.
LI
<cr>
After entering the command, you will be prompted to enter new values
one at a time. Press ENTER without giving a value to move to the next
parameter.
Example (shows default values):
>li
Cp offset : 0 ?
Cp gain : 1 ?
RH offset : 0 ?
RH gain : 1 ?
T offset : 0 ?
T gain : 1 ?
>