Vaisala Drycap DMT242 User Manual Download Page 1

 

 

Vaisala DRYCAP

®

 Dewpoint Transmitter DMT242

 

 

 

U

SER

'

G

UIDE 

 

June

 

2008

 

|

 

M010036EN-D 

INTRODUCTION 

DMT242 is a dewpoint transmitter for wide range of OEM applications. DMT242 measures dewpoint 
with excellent long term stability which is maintained automatically with the patented auto-calibration 
procedure. The Vaisala DRYCAP

®

 polymer sensor technology used in DMT242 is also durable against 

dew in case condenced water exists in the process during system malfunction. The product mechanics 
have been designed for harsh environments requiring protection against dust, dirt and splashed water. 
The disconnection and reconnection of the transmitter is easy with the connector where the output signal 
and supply voltage wires are connected. The unit also has a serial line for rescaling the analog output. 

PRESSURE SETTING PROCEDURE FOR PRESSURIZED PROCESSES 

 

For achieving the most accurate measurements 
in pressurized processes, set the process 
pressure to DMT242 according to Figure 1 by 
using the pressure switches (see Figure 3, item 
8). As shipped from factory, the pressure 
switch setting is 1 bar, as in switch number 4 in 
the ON position.

 

MOUNTING 

1.

 

Insert the sealing washer (see Figure 1) on 
the probe and set the probe through the 
fitting of the process pipe. The probe has 
G½" ISO 228/1 parallel thread. 

2.

 

Fasten the transmitter to the fitting of the 
process pipe by tightening from the nut of 
the probe (24 mm). 

WIRING 

1.

 

Remove the cover. 

2.

 

Take out the connector. 

3.

 

Take out the screw terminal from the 
connector by pushing it out, for example, 
with the fixing screw. 

4.

 

Use a three-wire cable. Suitable 2 m or 
10 m optional cable is available from 
Vaisala (items: 221475 for 2 m cable and 
221476 for 10 m cable). Connect the wires 
to the connector terminals as follows: 

 

2 3 4

1

Switches

Pressure (absolute)

2 … 4

0 … 2

12 … 14

4 … 6

8 … 10

10 … 12

14 … 16

6 … 8

18 … 20

16 … 18

(bar)

 

Figure 1 Pressure Setting Table 

 

Figure 2 DMT242 

 

Terminal nr 1 = V (VAC line) 
Terminal nr 2 = V supply - (VAC neutral)/signal - 
Terminal nr 3 =  
Leave the ground terminal free

 

+

-

THE SWITCH MARKED WITH 
A BLACK SQUARE IS IN 
THE

 ON 

POSITION

 

1

 

=   Cover 

2

 

=   Connector 

3

 

=   Flat gasket 

4

 

=   Counter connector 

5

 

=   Transmitter body 

6

 

=   Nut 

7

 

=   Sealing washer 

8

 

=   Probe 

Wire colours in cables 
221475 and 221476: 

1 = brown 
2 = blue  
3 = black 