5.6.2 Stop Measurement Output
You can stop the measurement output by pressing the
Esc
key, or with the
S
command:
S
<cr>
5.6.3 Set Output Interval
Use the
INTV
command to change the output interval of the automatically repeating
measurement messages. The measurement messages are repeated in the
RUN
mode, or after
the
R
command has been given.
Table 28 INTV Command
Syntax
Description
INTV [n xxx]
<cr>
Set the output interval:
n
time interval = 0 … 255
xxx
time unit =
S
,
MIN
, or
H
Example:
intv 1 min
Output interval: 1 MIN
The shortest output interval (with n = 0) outputs the measurement messages as quickly as the
transmitter’s internal measurement cycle produces them, without additional delay.
5.6.4 Output a Reading Once
Use the
SEND
command to output a single measurement message. The output message is sent
in the format that has been set using the
FORM
command.
Syntax
Description
SEND
[address]
<cr>
[address]
= address of the transmitter, range 0 … 99.
Must be specified if the transmitter is in
POLL
mode, and a line
has not been opened using the
OPEN
command.
Example:
send
Tdf= -2.0 'C H2O= 5152 ppm
Chapter 5 – Serial Line Operation
47