![Vaisala Drycap DMT152 User Manual Download Page 19](http://html1.mh-extra.com/html/vaisala/drycap-dmt152/drycap-dmt152_user-manual_3228362019.webp)
4.3.1 Output Noise Minimization Option
To minimize the output noise:
• Wire only the analog outputs (V/mA) to Port I
• Connect the power supply only to Port II
Avoid connecting the Port I and Port II grounds (pin #3) together
1
3
4
2
1
3
4
2
3
+
-
V/mA
4
+
-
V/mA
2
+
-
Vs
1
3
+ / A
4
2
RS485
- / B
Figure 5 Output Noise Minimization Wiring Option
The wiring of pinouts 1 … 4 in ports I and II in the output minimization option is described in
the following table.
Table 5 Output Noise Minimization Wiring Option Pinouts and Wire Colors
Pin
Connector I
Connector II
Wire Color
1
Not wired
VDC
Brown
2
Signal Ch 1+
RS-485 - / B
White
3
GND
GND
Blue
4
Signal Ch 2+ / LED
RS-485 + / A
Black