![Vaisala DPT146 User Manual Download Page 47](http://html1.mh-extra.com/html/vaisala/dpt146/dpt146_user-manual_3221748047.webp)
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 47
Set Output Interval
Use the
INTV
command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the
R
command has
been given.
INTV
[
n xxx
]<cr>
where
n
= time interval, range 0 ... 255
xxx
= time unit = "S", "MIN", or "H"
The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the transmitter’s internal measurement cycle
produces them, without additional delay.
Example:
intv 1 min
Output interval: 1 MIN
Output a Reading Once
Use the
SEND
command to output a single measurement message. The
output message is sent in the format that has been set using the
FORM
command.
SEND
[
address
]<cr>
where
address = Address of the transmitter, range 0 ... 255.
Specifying the address is only necessary if the target
transmitter is in POLL mode, and has not been accessed
with the
OPEN
command.
Example:
send
Tdf= 12.5 'C P= 0.991 bara T= 24.4 'C H2O= 15504 ppm
Tdfatm= 13.5 'C