![Vaisala DPT146 User Manual Download Page 46](http://html1.mh-extra.com/html/vaisala/dpt146/dpt146_user-manual_3221748046.webp)
USER'S GUIDE____________________________________________________________________
46 __________________________________________________________________ M211372EN-B
Serial Line Output Commands
Start Measurement Output
Use the
R
command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the
FORM
command.
R
<cr>
Example (measurement message in default format):
r
Tdf= 12.5 'C P= 0.990 bara T= 24.4 'C H2O= 15489 ppm
Tdfatm= 13.5 'C
Tdf= 12.5 'C P= 0.990 bara T= 24.4 'C H2O= 15488 ppm
Tdfatm= 13.5 'C
Tdf= 12.5 'C P= 0.990 bara T= 24.4 'C H2O= 15502 ppm
Tdfatm= 13.5 'C
...
Outputting the results continues in intervals issued with the command
INTV
. You can stop the output by pressing the
Esc
key, or with the
command
S
. Since the interface is half-duplex, you must enter the
commands when the transmitter is not outputting.
Stop Measurement Output
You can stop the measurement output by pressing
ESC
, or with the
S
command:
S
<cr>