Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 45
Set Serial Line Settings
Use the
SERI
command to set the serial line settings. The new
settings will be taken into use when the transmitter is reset or
powered up.
Note that changing the serial line settings will slightly alter how the
transmitter works with the MI70 indicator. If you are not using the
default settings of 19200 N 8 1, you will have to power off the MI70
before it can find the transmitter (you will not be able to connect it
on-the-fly).
SERI
[
b p d s
]<cr>
where
b
= baud rate (300, 600, 1200, 2400, 4800, 9600, or 19200)
p
= parity (n = none, e = even, o = odd)
d
= data bits (7 or 8)
s
= stop bits (1 or 2)
Example:
seri 9600 e 7 1
Baud P D S : 9600 E 7 1
Set Transmitter Address
To operate in the POLL mode, the transmitter must have an address.
If multiple transmitters share the same serial line, each transmitter
must have a different address.
For a description of the serial interface modes, see section Set Serial
Line Operating Mode on page 44.
ADDR
[
address
]<cr>
where
address
= Transmitter address, range 0 ... 255.
Example:
addr
Address : 0 ?