![vacuumbrand PC 8 / MD 12C Instructions For Use Manual Download Page 27](http://html1.mh-extra.com/html/vacuumbrand/pc-8-md-12c/pc-8-md-12c_instructions-for-use-manual_3132834027.webp)
Installing a pump in a vacuum system
➨
If dangerous or polluting fluids could be released at the
outlet, install an appropriate system to catch and dis-
pose of those fluids.
+
Connect a gas-tight exhaust line at the pump outlet if
necessary. Always vent exhaust gases appropriately
(e.g., into a fume hood).
+
Never block the gas outlet. The exhaust line must al-
ways be free of obstructions (no back pressure) to en-
sure an unimpeded discharge of gas. The cross-sec-
tion of the outlet tubing must be at least the size of the
pump’s exhaust connection.
+
Particles and dust must not be aspirated. If necessary,
you must install appropriate filters. You must ensure
their suitability concerning gas flow, chemical resis
-
tance and resistance to clogging prior to use.
+
Make sure ventilation is adequate, especially if the pump
is installed in an enclosure, or if the ambient temperature
is elevated. Provide external ventilation, if necessary.
• Reduce the transmission of vibration. Prevent mechan
-
ical load due to rigid pipelines. Insert elastic hoses or
flexible elements as couplings between the pump and
rigid pipes.
Note
: Flexible elements will compress or flatten when
evacuated if not designed for use under vacuum.
• Hose connections at the pump inlet must always be
gas tight.
• A power failure may cause accidental ventilation of the
pump, especially if the gas ballast valve is open . If this
constitutes a potential source of danger, take appropri-
ate safety measures.
Use and operation
Summary of Contents for PC 8 / MD 12C
Page 50: ...page 50 of 52...
Page 51: ...page 51 of 52...