60
20901224_EN_VACUU·PURE 10C_V1.8_221122
Error messages
6.2
6.2 Error indication
Error indication
Flashing
Description
1x
Temperature in the inadmissible range
2x
not assigned
3x
Motor current consumption in faulty range or
other motor error
4x
Fan defective
5x
Overvoltage or undervoltage in the intermediate
circuit of the frequency converter
6x
Error / communication interruption VACUU·BUS
accessories
7x
Other errors (software versions control board
/ frequency converter incompatible with each
other, other frequency converter errors)
>
If warnings and errors occur simultaneously, only the errors
(red LED) are displayed.
>
Several errors are indicated one after the other by combined
flashing cycles.
>
An error message is displayed until it is acknowledged.
Acknowledge an error message by switching the power
switch off and on after you have eliminated the error.
Example
Error
LED flashing rate
Temperature errors (1x)
and other errors (7x) occur
at the same time
1x
1x
7x
7x
Which and how many errors are present can be identified from
the flashing cycles.
Example
Pending error
Possible flashing
rates in the event of
an error