47
20901224_EN_VACUU·PURE 10C_V1.8_221122
Commissioning (operation)
>
The vacuum pump does not create a vacuum‑tight seal when
it switches off.
>
If necessary, install an in‑line solenoid valve or shut‑off valve
in the inlet line to isolate your application from the vacuum
pump with a vacuum‑tight seal.
5.2.2
5.2.2 Regeneration mode
Regeneration mode
The regeneration mode is used to quickly dry the pump interior
after the application is finished or before it is taken out of service.
Here, ambient air is fed into the interior of the pump and the
interior is dried by the air supply.
The pump does not need to be separated from the application
for regeneration
.
The pump runs at a reduced speed during regeneration.
The air for regeneration mode is drawn in via a filter on the rear
of the vacuum pump. The ambient air is aspirated here.
>
Check the filter regularly for dirt and blockages.
>
Replace dirty or clogged filters,
see chapter:
7.4 Air inlet filter on page 76
7.4 Air inlet filter on page 76
Start regeneration mode
5
The vacuum pump runs at a reduced speed and aspirates
ambient air.
5
The inside of the pump is dried.
5
Regeneration mode ends automatically after a period of
one hour.
Start regeneration
mode
IMPORTANT!
Press key
Drying
(regeneration) with
ambient air
Summary of Contents for VACUU-PURE 10C
Page 1: ...Instructions for use OI no 20901224 Original instructions EN VACUU PURE 10C Vacuum pump ...
Page 12: ...12 20901224_EN_VACUU PURE 10C_V1 8_221122 Introduction ...
Page 93: ...93 20901224_EN_VACUU PURE 10C_V1 8_221122 ...
Page 94: ...94 20901224_EN_VACUU PURE 10C_V1 8_221122 ...