Installation and connection
Connect the outlet line
>
Remove the blind flange on the outlet flange.
>
Connect an outlet line with small flange KF DN 25 to the
outlet flange ensuring the connection is gas‑tight.
>
Alternatively, you can use an adapter from small flange
KF DN 25 on the hose nozzle and attach the outlet line to
it. Use an outlet line with an interior diameter of at least
19 mm,
see chapter:
8.2 Ordering information on page
8.2 Ordering information on page
. Secure hose connections on hose nozzles, e.g., with
a hose clip.
>
Route the outlet line with a fall from the outlet, i.e., running
downward so that no backup forms.
>
The length of the outlet line must not exceed 5 m. An
excessively long outlet line can lead to inadmissibly high
back pressure at the outlet and impair the function of the
sealing gas.
Connect the vapor condenser (EK) (option)
>
The vapor condenser requires increased ground clearance.
Mount the vacuum pump e.g., on the
VACUU·PURE shuttle
VACUU·PURE shuttle
before connecting the vapor condenser
see chapter:
Ordering information on page 85
Ordering information on page 85
.
>
Connect the vapor condenser with small flange KF DN 25 to
the outlet flange, ensuring the connection is gas‑tight.
>
Fasten the separator flask using the joint clamp.
IMPORTANT!
Outlet line at the
outlet
Connect the vapor
condenser
Summary of Contents for VACUU-PURE 10C
Page 1: ...Instructions for use OI no 20901224 Original instructions EN VACUU PURE 10C Vacuum pump ...
Page 12: ...12 20901224_EN_VACUU PURE 10C_V1 8_221122 Introduction ...
Page 93: ...93 20901224_EN_VACUU PURE 10C_V1 8_221122 ...
Page 94: ...94 20901224_EN_VACUU PURE 10C_V1 8_221122 ...