Prevent the discharge of dangerous gases and vapors
to the surrounding atmosphere. If appropriate, connect
the exhaust line to a suitable treatment system.
+
Never block the gas outlet ((8) hose nozzle for tubing
I.D. 3/8” (10 mm)). The exhaust hose has always to be
unobstructed and without back pressure to enable an
unhindered discharge of gases and protect the pump
valves from damage.
•
Note
: Install the hoses of the cooling system in a way
to avoid the flow / dripping of condensed water onto the
pumping unit (especially cables and electronic parts,
see also IP degree of protection, „Technical data“, pg.
• Ensure that the
coolant outlet tubing
is always unob-
structed and that it cannot get blocked.
• Maximum permissible coolant pressure at the exhaust
waste vapor condenser: 87 psi (6 bar) absolute. Outlet
flow must always be unhindered.
• Comply with the maximum permissible coolant pres
-
sures of additional components in the coolant circuit
(e.g., coolant valve).
• Avoid overpressure in the coolant circuit (e.g., caused
by blocked or squeezed coolant hoses).
• Only install the optional coolant valve in the supply line
of the exhaust waste vapor condenser.
Notes regarding the (dry) ice condenser
(PC 3001 TE VARIO
pro
)
The
exhaust waste vapor condenser
en-
ables an efficient condensation of the pumped
vapors at the outlet.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass
splinters in case of breaking, acts as ther-
mal insulation to avoid the condensation of
humidity and is intended to absorb shocks.
outlet (gas!)
hose nozzle