page 39 of 70
Replacing diaphragms and valves
+
Please read section ”Replacing diaphragms and valves” com-
pletely before starting maintenance.
The pictures may show other versions of pumps. This does not change
the method of replacing diaphragms and valves.
➨
Never operate the pump if covers or other parts of
the pump are disassembled.
➨
Before starting maintenance,
disconnect the electri-
cal power cord
.
Wait five seconds
after isolating the
equipment from AC power to allow the capacitors to
discharge.
➨
Ensure that the pump cannot be operated accidentally.
➨
Note
: The pump might be contaminated with the pro-
cess chemicals that have been pumped during opera-
tion. Ensure that the pump is decontaminated before
maintenance.
➨
Avoid the release of pollutants.
+
Never operate a defective or damaged pump
.
+
Check every motor capacitor regularly by measuring
its capacity and estimating its service life. Replace old
capacitors early enough to prevent a failure. The ca-
pacitors must be replaced by a trained electrician.
+
Take adequate precautions to protect people
from the effects of dangerous substances
that may have contaminated the pump and
may be released upon disassembly. Ensure
that the maintenance technician is famil-
iar with the safety procedures which relate
to the products processed by the pumping
system.
Use appropriate protective clothing, safety
goggles and protective gloves.
Summary of Contents for ME 16 NT
Page 42: ...page 42 of 70 3 4 5 4 4x 4 4x 4 2x...
Page 43: ...page 43 of 70 Don t loosen this screw 7 8 6 5 2x...
Page 44: ...page 44 of 70 9 10 11 5 2x...
Page 46: ...page 46 of 70 12 2 5 mm 13 14 2 5 mm 5 8x...
Page 47: ...page 47 of 70 15 696819 696827 16 17 Clean...
Page 48: ...page 48 of 70 696819 696827 18 19...
Page 50: ...page 50 of 70 Assembling the pump head 23 24 25 Pay attention to number of washers...
Page 52: ...page 52 of 70 29 30 31...
Page 54: ...page 54 of 70 35 Maintain all pump heads in the same way ME 16 NT MD 12 NT MV 10 NT 36A 36B...
Page 56: ...page 56 of 70 4 5 6 696827...
Page 57: ...page 57 of 70 7 8 size 2 2x 9 size 2 2x...
Page 59: ...page 59 of 70 Assembling the pump housing 37 38 39 5 2x 8 9 ft lbf 12 Nm...
Page 60: ...page 60 of 70 40 42 41 5 2x 8 9 ft lbf 12 Nm...
Page 61: ...page 61 of 70 45 4 2x 3 7 ft lbf 5 Nm 44 4 4x 3 7 ft lbf 5 Nm 43...
Page 62: ...page 62 of 70 46 4 4x 3 7 ft lbf 5 Nm 47 4 4x 3 7 ft lbf 5 Nm 48...