General view basic setup VACUU•LAN
setting process
parameters
➨
lower pressure value (condition for automatic shut down)
➨
upper pressure value (condition for restart)
➨
time for automatic shut down
setting the basic setup
basic setup
VACUU•LAN
start: pump down
☞
☞
☞
☞
☞
The pump (and the coolant valve if applicable) is switched
off if the pressure is below the lower pressure value for
longer than the time for automatic shut down.
☞
☞
☞
☞
☞
The pump starts automatically if the pressure is higher
than the upper set point or at great pressure increase.
connect and
configurate
cooling
water valve
yes
no
connect and
configurate
venting valve
yes
no
switching coolant
valve?
switching venting
valve?