Setting up and installing the equipment

  Equipment must be connected only to a 

suitable electrical supply

 and a suit-

able ground point. As such, the plug must be plugged into an outlet that is proper-

ly grounded. Failure to connect the motor to ground may result in deadly electrical 

shock.

  The supply cable may be fitted with a molded European IEC plug or a plug suit-

able for your local electrical supply. The cable contains wires color coded as fol-

lows: green or green and yellow: ground; blue or white: neutral; brown or black: 

hot.

+

  Due to the high compression ratio, the pump may generate overpressure at the 

outlet. Check pressure compatibility with system components (e.g., exhaust pipe-

line or exhaust valve) at the outlet.

+

  Do not permit any 

uncontrolled pressurizing

. Make sure that the exhaust tub-

ing cannot become blocked to avoid a 

risk of bursting

+

  Keep the electrical power cord away from heated surfaces.

•  Provide a firm, level platform for the equipment. Check that the system which you 

are going to evacuate is mechanically stable. Check that all fittings are secure. 

Ensure a stable position of the pumping unit.

•  Comply with 

maximum permissible pressures

 at inlet and outlet and with 

max-

imum permissible pressure differences

 between inlet and outlet. See section 

“Technical data“. Do not operate the pump with overpressure at the inlet.

•  Avoid overpressure of more than 1.2 bar absolute in the event that inert gas is 

connected to the pump, to the gas ballast or to a venting valve.

• 

Note

: Flexible elements will shrink when evacuated.

•  Connect hoses gas tight at inlet and outlet of the pump.

•   Ensure that no foreign objects can be drawn into the pump.

•   Check the power source and the pump’s rating plate to be sure that the power 

source and the equipment match in voltage, phase, and frequency.

•   Use only oil of the recommended type. Other oils or operating fluids may cause 

damage of the pump or danger. 

 Use

 special oils

 for the rotary vane pump, if, e.g.,

  -  the pump operates in the vicinity of potential ignition sources.

  -  oxygen or other flammable gases account for a large proportion of the evacu-

    ated gases.

•   Under normal operation conditions, pump oils and lubricants are not toxic and 

their use entails no danger. Certain hazards are, however, associated with some 

of these products. 

 •  Take adequate precautions when handling pump fluids, lubricants, and solvents. 

Use appropriate protective clothing, safety goggles and protective gloves to avoid 

excessive contact with the skin and possible skin irritations (including dermatitis). 

Do not inhale or swallow. Maintain adequate levels of hygiene and cleanliness. 

Ensure that the pump location is well ventilated and that possible toxic effects of 

certain vapors are avoided.

•  Comply with all relevant statutory requirements and regulations concerning the 

handling, storage and disposal of oil.

 

Attention

! Do not allow oils to be poured into or enter the drainage system or 

other bodies of water.

  Spillage can cause accidents, use suitable means of removing spilled oil.

Summary of Contents for HP 40B2

Page 1: ...completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999069 28 04 2016 Technology for Vacuum Systems...

Page 2: ...sted intensively including an endurance run of 24 hours Any faults even those which occur rarely are identified and can be eliminated immediately After completion of the endurance run every pumping un...

Page 3: ...and operation 14 Setting up and installing 14 Getting started 14 During operation 15 Shutdown 16 Accessories 16 Troubleshooting 16 Maintenance 16 Repair Maintenance Return Calibration 17 EC Declaratio...

Page 4: ...immediately Store the equipment in dry and non corrosive conditions Intended use Do not use the pump or any system parts on humans or animals Ensure that the individual components are only connected c...

Page 5: ...Do not operate the pump with overpressure at the inlet Avoid overpressure of more than 1 2 bar absolute in the event that inert gas is connected to the pump to the gas ballast or to a venting valve N...

Page 6: ...ons and guidelines Implement the required actions and adopt suitable safety mea sures Ambient conditions Adopt suitable measures in case of differences from recommended conditions e g using the equipm...

Page 7: ...bstances from the pump outlet install an appropriate collecting and disposal system and take protective action for pump and environment You must take suitable precautions to prevent any formation of e...

Page 8: ...ssary Attention If the maximum permissible inlet pressure or the backing pressure at the diffusion pump are exceeded pump fluid might enter other parts of the vacuum system Provide appropriate protect...

Page 9: ...Use appropriate protective clothing safe ty goggles and protective gloves Wear parts have to be replaced regularly Never operate a defective or damaged pump Vent the pumping unit before starting maint...

Page 10: ...ump Pump fluid capacity of diffusion pump ml 30 Maximum pumping speed l s 22 37 Ultimate vacuum mbar 1 x 10 6 Maximum permissible inlet outlet pressure absolute bar 1 1 Maximum pressure difference bet...

Page 11: ...ffusion pump AX 65 2 Backing pump RZ 2 5 3 Backing pump RZ 6 4 ON OFF switch diffusion pump 5 ON OFF switch backing pump 6 Rating plate of pumping unit 7 Connection diagram 8 Handle Position Component...

Page 12: ...idity of this document with respect to his product Manual no 999069 28 04 2016 An oil mist filter for installation at the outlet of the rotary vane pump RZ 6 is separately packed Pay attention to the...

Page 13: ...be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999069 28 04 2016 HP 63B...

Page 14: ...matic ventilation system Use connecting hoses with large diameter and keep them as short as possible to avoid flow losses Locate the pump as closely as possible to the application Always install outle...

Page 15: ...V2 once the pressure inside the vacuum vessel is below 0 1 mbar Open fore vacuum valve V1 Open high vacuum valve V3 Note If the fore vacuum M1 rises above 0 1 mbar close the high vacuum valve V3 and p...

Page 16: ...e diffusion pump Proceeding in accordance with the instructions given above ensures that the diffu sion pump is kept under vacuum no condensate build up in the pump fluid To avoid possible formation o...

Page 17: ...t are no original components Drain the device completely of fluids and residues Decontaminate the device Close all openings airtight especially if using substances hazardous to health To expedite repa...

Page 18: ...ed for service staff 11 If the paint is damaged we wish a repaint or a replacement of parts for reason of appearance repaint and replacement at customer s expense yes no 12 Legally binding declaration...

Page 19: ...de pompage Typ Type Type HP 40B2 HP 63B2 Artikelnummer Order number Num ro d article 699029 2612089 699037 Seriennummer Serial number Num ro de s rie Siehe Typenschild See rating plate Voir plaque si...

Page 20: ...lts obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of...

Reviews: