Use and operation

 Setting up and installing

+

 Make sure ventilation is adequate, especially if the pump is installed in an en-

closure, or if the ambient temperature is elevated. Provide external ventilation, if 

necessary.

•  Reduce the transmission of vibration. Prevent mechanical load due to rigid pipe-

lines. Insert elastic hoses or flexible elements as couplings between the pump 

and rigid pipes. 

 

Note

: Flexible elements will compress or flatten when evacuated if not designed 

for use under vacuum.

•  Hose connections at the pump inlet must always be gas tight.

•  If the pump is installed at an altitude of more than 2000 m above mean sea level, 

check compatibility with applicable safety requirements, and adopt suitable mea-

sures. There is a risk of the motor overheating due to insufficient cooling.

Keep a minimum distance of 5 cm between cooling fans and surrounding items (e.g., 

housing, walls, etc.), or else install an external automatic ventilation system.

Use connecting hoses with large diameter and keep them as short as possible to 

avoid flow losses. Locate the pump as closely as possible to the application.

Always install outlet tubing descending from the pump or provide other measures to 

avoid backflow of condensate towards the pump. 

NOTICE

Proceed strictly according to the given order and comply with the given time 

and pressure data.

Switching on the pumping unit

 

(pump and vacuum vessel at atmospheric pressure)

  Open fore-vacuum valve (V1). Close by-pass valve (V2), high vacuum valve (V3), 

and venting valve (V4).

➨ 

Switch on backing pump (rotary vane pump RZ 2.5 or RZ 6), switch on diffusion 

pump.

The diffusion pump can be switched on only once the backing pump has been 

switched on.

➨ 

Close fore-vacuum valve (V1) after the heating time (approx. 15 minutes) has 

elapsed and the fore-vacuum (M1) is below 0.1 mbar.

➨ 

Open by-pass valve (V2).

➨ 

Close by-pass valve (V2) once the pressure inside the vacuum vessel is below 

0.1 mbar.

➨ 

Open fore-vacuum valve (V1).

➨ 

Open high vacuum valve (V3).

+

 

Note:

 If the fore-vacuum (M1) rises above 0.1 mbar, close the high vacuum valve 

(V3) and pump down with the backing pump to a pressure below 0.1 mbar.

+

 

If the thermal cutout of the rotary vane pump has tripped, it is imperative 

that the high vacuum valve (V3) be closed and the diffusion pump switched 

off!

NOTICE

 Getting started

Summary of Contents for HP 40B2

Page 1: ...completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999069 28 04 2016 Technology for Vacuum Systems...

Page 2: ...sted intensively including an endurance run of 24 hours Any faults even those which occur rarely are identified and can be eliminated immediately After completion of the endurance run every pumping un...

Page 3: ...and operation 14 Setting up and installing 14 Getting started 14 During operation 15 Shutdown 16 Accessories 16 Troubleshooting 16 Maintenance 16 Repair Maintenance Return Calibration 17 EC Declaratio...

Page 4: ...immediately Store the equipment in dry and non corrosive conditions Intended use Do not use the pump or any system parts on humans or animals Ensure that the individual components are only connected c...

Page 5: ...Do not operate the pump with overpressure at the inlet Avoid overpressure of more than 1 2 bar absolute in the event that inert gas is connected to the pump to the gas ballast or to a venting valve N...

Page 6: ...ons and guidelines Implement the required actions and adopt suitable safety mea sures Ambient conditions Adopt suitable measures in case of differences from recommended conditions e g using the equipm...

Page 7: ...bstances from the pump outlet install an appropriate collecting and disposal system and take protective action for pump and environment You must take suitable precautions to prevent any formation of e...

Page 8: ...ssary Attention If the maximum permissible inlet pressure or the backing pressure at the diffusion pump are exceeded pump fluid might enter other parts of the vacuum system Provide appropriate protect...

Page 9: ...Use appropriate protective clothing safe ty goggles and protective gloves Wear parts have to be replaced regularly Never operate a defective or damaged pump Vent the pumping unit before starting maint...

Page 10: ...ump Pump fluid capacity of diffusion pump ml 30 Maximum pumping speed l s 22 37 Ultimate vacuum mbar 1 x 10 6 Maximum permissible inlet outlet pressure absolute bar 1 1 Maximum pressure difference bet...

Page 11: ...ffusion pump AX 65 2 Backing pump RZ 2 5 3 Backing pump RZ 6 4 ON OFF switch diffusion pump 5 ON OFF switch backing pump 6 Rating plate of pumping unit 7 Connection diagram 8 Handle Position Component...

Page 12: ...idity of this document with respect to his product Manual no 999069 28 04 2016 An oil mist filter for installation at the outlet of the rotary vane pump RZ 6 is separately packed Pay attention to the...

Page 13: ...be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999069 28 04 2016 HP 63B...

Page 14: ...matic ventilation system Use connecting hoses with large diameter and keep them as short as possible to avoid flow losses Locate the pump as closely as possible to the application Always install outle...

Page 15: ...V2 once the pressure inside the vacuum vessel is below 0 1 mbar Open fore vacuum valve V1 Open high vacuum valve V3 Note If the fore vacuum M1 rises above 0 1 mbar close the high vacuum valve V3 and p...

Page 16: ...e diffusion pump Proceeding in accordance with the instructions given above ensures that the diffu sion pump is kept under vacuum no condensate build up in the pump fluid To avoid possible formation o...

Page 17: ...t are no original components Drain the device completely of fluids and residues Decontaminate the device Close all openings airtight especially if using substances hazardous to health To expedite repa...

Page 18: ...ed for service staff 11 If the paint is damaged we wish a repaint or a replacement of parts for reason of appearance repaint and replacement at customer s expense yes no 12 Legally binding declaration...

Page 19: ...de pompage Typ Type Type HP 40B2 HP 63B2 Artikelnummer Order number Num ro d article 699029 2612089 699037 Seriennummer Serial number Num ro de s rie Siehe Typenschild See rating plate Voir plaque si...

Page 20: ...lts obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of...

Reviews: