Technical data

Type

HP 40B2 /  

RZ 2.5

HP 40B2 / 

RZ 6

HP 63B2

Backing pump

RZ 2.5

RZ 6

Recommended oil for backing pump

VACUUBRAND B-oil

Diffusion pump

AX 65

Recommended pump fluid for 

diffusion pump

diffusion pump oil for

VACUUBRAND oil diffusion pump

Pump fluid capacity of diffusion pump

ml

30

Maximum pumping speed

l/s

22

37

Ultimate vacuum* 

mbar

< 1 x 10

-6

Maximum permissible inlet / outlet 

pressure (absolute)

bar

1.1

Maximum pressure difference 

between inlet and outlet 

bar

1.1

Maximum permissible pressure 

(absolute) at gas ballast valve

bar

1.2

Maximum permissible temperature 

of gaseous media

continuous operation: 40°C, 

for short periods (less than 5 minutes): up to 80°C

Permissible ambient temperature 

storage / operation

°C

-10 to +60 / +12 to +40

Permissible relative atmospheric 

moisture during operation 

%

30 to 85 (no condensation)

Maximum permissible installation 

altitude above mean sea level

m

2000

Maximum permissible range of 

supply voltage

230V~ 

±

10% 50/60Hz

Rated input current:

230 V~ 50/60 Hz

A

2.7 / 2.2

3.2 / 3.2

Power draw

230 V~ 50/60 Hz

VA

621 / 506

736 / 736

Degree of protection IEC 529

IP 10B

Inlet

small flange KF DN 40

clamping flange

ISO-K DN 63

Outlet

hose nozzle DN 10 mm

A-weighted emission sound pressure 

level** (uncertainty K

pA

: 3 dB(A))

dB(A)

42

49

Dimensions L x W x H  approx.

mm

445 x 385 x 435

445 x 381 x 460

Weight approx.

kg

25.4

30.9

Summary of Contents for HP 40B2

Page 1: ...completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999069 28 04 2016 Technology for Vacuum Systems...

Page 2: ...sted intensively including an endurance run of 24 hours Any faults even those which occur rarely are identified and can be eliminated immediately After completion of the endurance run every pumping un...

Page 3: ...and operation 14 Setting up and installing 14 Getting started 14 During operation 15 Shutdown 16 Accessories 16 Troubleshooting 16 Maintenance 16 Repair Maintenance Return Calibration 17 EC Declaratio...

Page 4: ...immediately Store the equipment in dry and non corrosive conditions Intended use Do not use the pump or any system parts on humans or animals Ensure that the individual components are only connected c...

Page 5: ...Do not operate the pump with overpressure at the inlet Avoid overpressure of more than 1 2 bar absolute in the event that inert gas is connected to the pump to the gas ballast or to a venting valve N...

Page 6: ...ons and guidelines Implement the required actions and adopt suitable safety mea sures Ambient conditions Adopt suitable measures in case of differences from recommended conditions e g using the equipm...

Page 7: ...bstances from the pump outlet install an appropriate collecting and disposal system and take protective action for pump and environment You must take suitable precautions to prevent any formation of e...

Page 8: ...ssary Attention If the maximum permissible inlet pressure or the backing pressure at the diffusion pump are exceeded pump fluid might enter other parts of the vacuum system Provide appropriate protect...

Page 9: ...Use appropriate protective clothing safe ty goggles and protective gloves Wear parts have to be replaced regularly Never operate a defective or damaged pump Vent the pumping unit before starting maint...

Page 10: ...ump Pump fluid capacity of diffusion pump ml 30 Maximum pumping speed l s 22 37 Ultimate vacuum mbar 1 x 10 6 Maximum permissible inlet outlet pressure absolute bar 1 1 Maximum pressure difference bet...

Page 11: ...ffusion pump AX 65 2 Backing pump RZ 2 5 3 Backing pump RZ 6 4 ON OFF switch diffusion pump 5 ON OFF switch backing pump 6 Rating plate of pumping unit 7 Connection diagram 8 Handle Position Component...

Page 12: ...idity of this document with respect to his product Manual no 999069 28 04 2016 An oil mist filter for installation at the outlet of the rotary vane pump RZ 6 is separately packed Pay attention to the...

Page 13: ...be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999069 28 04 2016 HP 63B...

Page 14: ...matic ventilation system Use connecting hoses with large diameter and keep them as short as possible to avoid flow losses Locate the pump as closely as possible to the application Always install outle...

Page 15: ...V2 once the pressure inside the vacuum vessel is below 0 1 mbar Open fore vacuum valve V1 Open high vacuum valve V3 Note If the fore vacuum M1 rises above 0 1 mbar close the high vacuum valve V3 and p...

Page 16: ...e diffusion pump Proceeding in accordance with the instructions given above ensures that the diffu sion pump is kept under vacuum no condensate build up in the pump fluid To avoid possible formation o...

Page 17: ...t are no original components Drain the device completely of fluids and residues Decontaminate the device Close all openings airtight especially if using substances hazardous to health To expedite repa...

Page 18: ...ed for service staff 11 If the paint is damaged we wish a repaint or a replacement of parts for reason of appearance repaint and replacement at customer s expense yes no 12 Legally binding declaration...

Page 19: ...de pompage Typ Type Type HP 40B2 HP 63B2 Artikelnummer Order number Num ro d article 699029 2612089 699037 Seriennummer Serial number Num ro de s rie Siehe Typenschild See rating plate Voir plaque si...

Page 20: ...lts obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of...

Reviews: