ColonyDoc-It Imaging Station
45
Print an Audit Trail (Image History)
Open the image for which an Audit
Trail print copy is needed.
Go to
Tools > Reports
. (This option
is disabled, if no printer is available.)
A window opens with various types
of reports available.
To select to print an Audit Trail, click
the
Image History
item. If other
reports are required, click in the
checkbox of the other reports
needed. For example, if both an
audit trail and the image report need
to be printed, click on
Image Report
and
Image
History
.
Adjust the header and footer settings
or printer settings if necessary, and
print the trail.
Servicing the ColonyDoc-It
Cleaning and Care
Use 70% Isopropanol Alcohol to clean the ColonyDoc-It external surfaces, base, and doors.
Clean the ColonyDoc-It twice a day, before and after every use.
Cleaning the Diopter
Weekly clean the diopter with a soft photographic lens cleaning cloth dampened with lens cleaning solution.
CAUTION:
Do not pour the solution directly onto the diopter.
The diopter is located in the filter box. To access the diopter:
Remove the four brass
thumbnuts on the filter box
and remove the filter box
from the housing.
Clean both sides of the
diopter. Remove the filter if
necessary to access the
diopter (refer to
Installing
the Filter
for instructions on
removing the filter.)