2.2
Air Requirements
The system pressure should be regulated between 100psi and
150psi
of
clean, dry, unregulated compressed air. Use of dryer system, air filters
and water trap is highly recommended.
Air Pressure
100 - 120 PSIG (585-825 kPA)
Amount of compressed air
approx.
8.5m
3
/h
Air Tubes Diameter
10
mm (3/8”)
2.2.1
Air Quality Requirements
The air supplied to the instrument should comply with ISO8573.1
Quality Class 3 as follows:
•
DIRT particle size:
< 5 microns
•
WATER Pressure Dew Point: <-4 F @ 100 PSIG
(128
ppm
by
volume)
•
OIL (including vapor):
< 1.0 ppm
If the air line will be subjected to temperatures that could promote
condensation, a water filter attached to the air line inside the lab is
recommended to remove any
condensation that may occur.