EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 14
Section 9.0 ASSEMBLY PRIOR TO USE
IMPORTANT
If at any time the unit performs erratically or provides otherwise abnormal operation,
remove the unit from service and have it inspected or repaired.
The CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR should be inspected upon
receipt and before each use. Damaged equipment should be removed from service
and returned to US Endoscopy for repair or replacement. Before each use, perform
the procedures and inspections described in Sections 9.1, 9.2 and 9.3
9.1 PREPARATION
1. Install the CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR on the Accessory
Cart that can be purchased separately, or on a flat surface, away from potential
sources of spraying or leaking liquids.
2. Visually inspect the CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR for
external signs of damage.
9.2 ELECTRICAL CONNECTIONS
1. Inspect the electrical connections. Do not use if inspection reveals any damage.
2. Connect the line cord (mains lead) to the AC Power Connection on the back of
the CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR.
3. Before connecting the line cord (mains lead) to the hospital grade wall outlet,
make sure that the main power switch is off and that the voltage is correct. The
CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR has a universal AC line input
device, the nominal AC line voltage is 100 to 240 VAC and the AC line
frequency is 50/60 Hz. The AC line voltage should not drop below 90 VAC or
exceed 264 VAC. There are no switches or other AC line configuration
requirements.